Free tool · CRCST exam topic
Steam sterilization cycle times
Pick the sterilizer type and what is in the load to see CDC’s minimum exposure time at 250°F (121°C) and 270°F (132°C), and the drying time.
Gravity displacement · Wrapped instruments
- Exposure at 270°F (132°C)
- 15 min
- Exposure at 250°F (121°C)
- 30 min
- Drying time
- 15–30 min
Gravity displacement · Textile packs
- Exposure at 270°F (132°C)
- 25 min
- Exposure at 250°F (121°C)
- 30 min
- Drying time
- 15 min
Gravity displacement · Wrapped utensils
- Exposure at 270°F (132°C)
- 15 min
- Exposure at 250°F (121°C)
- 30 min
- Drying time
- 15–30 min
Dynamic-air-removal (prevacuum) · Wrapped instruments
- Exposure at 270°F (132°C)
- 4 min
- Exposure at 250°F (121°C)
- Not applicable
- Drying time
- 20–30 min
Dynamic-air-removal (prevacuum) · Textile packs
- Exposure at 270°F (132°C)
- 4 min
- Exposure at 250°F (121°C)
- Not applicable
- Drying time
- 5–20 min
Dynamic-air-removal (prevacuum) · Wrapped utensils
- Exposure at 270°F (132°C)
- 4 min
- Exposure at 250°F (121°C)
- Not applicable
- Drying time
- 20 min
Minimum cycle times for wrapped items
CDC’s Table 7, adapted from AAMI standards, lists these minimum exposure and drying times:
| Sterilizer | Item | 250°F (121°C) | 270°F (132°C) | Drying |
|---|---|---|---|---|
| Gravity displacement | Wrapped instruments | 30 min | 15 min | 15–30 min |
| Gravity displacement | Textile packs | 30 min | 25 min | 15 min |
| Gravity displacement | Wrapped utensils | 30 min | 15 min | 15–30 min |
| Dynamic-air-removal (prevacuum) | Wrapped instruments | Not applicable | 4 min | 20–30 min |
| Dynamic-air-removal (prevacuum) | Textile packs | Not applicable | 4 min | 5–20 min |
| Dynamic-air-removal (prevacuum) | Wrapped utensils | Not applicable | 4 min | 20 min |
Immediate-use steam sterilization (IUSS)
CDC’s Table 8 gives examples of cycles at 270°F (132°C) for unwrapped items used immediately:
| Sterilizer | Load | Time at 132°C |
|---|---|---|
| Gravity displacement | Nonporous items only | 3 min |
| Gravity displacement | Mixed nonporous and porous items | 10 min |
| Prevacuum (dynamic-air-removal) | Nonporous items only | 3 min |
| Prevacuum (dynamic-air-removal) | Mixed nonporous and porous items | 4 min |
| Steam-flush pressure-pulse | Nonporous or mixed items | 4 min (and the manufacturer’s instructions) |
The four parameters of steam sterilization
CDC names four: steam, pressure, temperature and time.
- Steam: ideally dry saturated steam that contacts every surface. Trapped air blocks that contact, which is why air removal matters.
- Pressure: raising the pressure is what lets steam reach sterilizing temperatures.
- Temperature: 121°C (250°F) and 132°C (270°F) are the two common set points.
- Time: the exposure time at temperature, which does not include come-up or drying time.
Each load is checked with mechanical, chemical and biological monitoring; see thesterilization indicators guide.
Steam sterilization on the CRCST exam
Sterilization Process is 21% of the CRCST exam in HSPA’s content outline. Cycle types, loading, cycle parameters, drying, wet packs and process failures all come up, alongside low-temperature methods such as ethylene oxide and hydrogen peroxide.
Test yourself: a CRCST-style question
According to CDC, what is the minimum exposure time for wrapped instruments in a dynamic-air-removal (prevacuum) steam sterilizer at 270°F (132°C)?
- 3 minutes
- 4 minutes
- 10 minutes
- 15 minutes
Show answer
B. 4 minutes
CDC Table 7 lists 4 minutes at 132°C for wrapped instruments in a prevacuum sterilizer, with 20–30 minutes of drying. A gravity displacement sterilizer needs 15 minutes at 132°C (or 30 minutes at 121°C).
Questions like this come up in the Sterilization Process section of the CRCST exam. The 901 questions in the app each come with a written explanation.
FAQ
- What are the two most common steam sterilization temperatures?
- 121°C (250°F) and 132°C (270°F). CDC gives recognized minimum exposure periods for wrapped supplies of 30 minutes at 121°C in a gravity displacement sterilizer, or 4 minutes at 132°C in a prevacuum sterilizer.
- Why are prevacuum cycles so much shorter?
- A dynamic-air-removal sterilizer pulls air out of the chamber and packs with a vacuum before steam is admitted, so steam reaches every surface quickly. A gravity sterilizer relies on steam pushing the air out through a drain, which is slower and less reliable, so exposure times are longer.
- Are these the times to set on our sterilizer?
- No. They are minimums from CDC’s guideline for exam study. Always use the cycle validated by the sterilizer manufacturer and the device’s instructions for use (IFU); many devices and containers specify extended cycles.
- What is immediate-use steam sterilization (IUSS)?
- A shorter cycle, formerly called “flash” sterilization, for items needed right away that cannot be packaged, sterilized and stored first. It is not meant for convenience or to replace adequate instrument inventory, and items should not be stored after IUSS.
Sources
- CDC. Table 7: Minimum cycle times for steam sterilization cycles
- CDC. Table 8: Examples of flash steam sterilization parameters
- CDC. Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization
- HSPA. CRCST Exam Content Outline (revised November 2023)
Checked against CDC on November 13, 2026. Minimum times from CDC’s 2008 guideline, for CRCST exam study only. Always follow the sterilizer manufacturer’s validated cycles and each device’s instructions for use (IFU), your facility’s policy and ANSI/AAMI ST79.